Certification Record
LC1227412
H2Scan Corporation
2023-08-17
California
United States
Barrier P/N 50000021, HY-ALERTA Model 1600 Series, HY-OPTIMA Model 1700 and 1700B Series
29.260.20
- IS Shunt Safety Assembly
- Model: Barrier P/N 50000021 (To be used to limit power to the HY-ALERTA Model 1600 Series and HYOPTIMA Model 1700 and 1700B Series to an intrinsically safe level)
- Ratings:
- [Ex ib] IIC Gb (CSA rating)
- Class I Zone 1 [AEx ib] IIC Gb (UL rating)
- Electrical parameters:
- Um = 250 V
- Uo = 11.5 Vdc
- Io = 3.27 Adc
- Po = 9.41 W
- Ci = 0
- Li = 0
- Co = 1.64uF
- Lo = 3.3uH
- Lo/Ro = 29 μH/ohms
- Process Monitor
- Model: HY-ALERTA Model 1600 Series & HY-OPTIMA Model 1700 and 1700B Series (to be used with associated IS Shunt Safety Assembly)
- Ratings:
- Input: 5 – 28Vdc (10Vdc nominal), 10W;
- Output: 4 – 20mA
For units with sensing element intended to be used in Zone 0:
– Sensor ratings:
- Ex ia II (H2) T3 Ga (CSA rating)
- Class I Zone 0 AEx ia II (H2) T3 Ga (UL rating)
- Class I Division 1 Group B T3 (CSA & UL rating)
– Analyzer ratings:
- Ex db+db IIB T3 Ga (CSA rating)
- Class I Zone 0 AEx db+db IIB T3 Ga (UL rating)
- Class I Division 1 Group CD T3 (CSA & UL rating)
For units with sensing element intended to be used in Zone 1
– Sensor ratings:
- Ex ib II (H2) T3 Gb (CSA rating)
- Class I Zone 1 AEx ib II (H2) T3 Gb (UL rating)
- Class I Division 2 Group B T3 (CSA & UL rating)
– Analyzer ratings:
- Ex db IIB T3 Gb (CSA rating)
- Class I Zone 1 AEx db IIB T3 Gb (UL rating)
- Class I Division 2 Group CD T3 (CSA & UL rating)
The maximum intrinsically safe input and output parameters:
- J1 Pins 2 & 3 (power supply)
- Ui = 11.5 V
- Ii = 3.27 A
- Pi = 9.41 W
The equivalent parameters are:
- CI = 0
- LI = 0
- J1 Pins 4 & 5 (4-20 mA loop)
- UI = 26.7 V
- Ii = 91 mA
- Pi = 0.611 W
The equivalent parameters are:
- CI = 68.2 nF
- LI = 0
- J2 Pins 1 & 2 and 3 & 4 (Rx and Tx)
- UI = 3.7 V
- Ii = 225 mA
- Pi = 206 mW
The equivalent parameters are:
- CI = 0
- LI = 0
- Uo = 5.9 V
- Io = 238 mA
- Po = 0.35 W
- Co = 1000 μF
- Lo = 2.5 mH
- Lo/Ro = 404 μH/ohms
IS Shunt Safety Assembly
- CAN/CSA C22.2 No. 60079-0:19 / UL 60079-0 (Ed. 7)
- CAN/CSA C22.2 No. 60079-11:14 (R2018) / UL 60079-11 (Ed. 6)
- CAN/CSA C22.2 No. 61010-1A-12 / UL 61010-1 (Ed. 3)
Process Monitor
- CAN/CSA C22.2 No. 60079-0:19 / UL 60079-0 (Ed. 7)
- CAN/CSA C22.2 No. 60079-1:16 / UL 60079-1 (Ed. 7)
- CAN/CSA C22.2 No. 60079-11:14 (R2018) / UL 60079-11 (Ed. 6)
- CAN/CSA C22.2 No. 60079-26:16 / UL 60079-26 (Ed. 3)
- CAN/CSA C22.2 No. 61010-1A-12 / UL 61010-1 (Ed. 3)
- None
General Note:
- Project no.: 18552/18552A
- LC Certificate no.: LC20CA18552-1
- LC Listing report no.: 11.00.9001-1_Rev.4
- Rev.1 – LC Listing report no. 9001-1S_Rev.1 under the project 9001
- Rev.2 – LC Certificate no. 9001-1S_Rev.2 under the project LV-14007
- Rev.3 – LC Certificate no. 9001-1S_Rev.3 under the project 12618
- Rev.4 – LC Certificate no. LC20CA18552-1 under the project 18552/18552A
– Refer to the listing report no. 11.00.9001-1_Rev 4.
– Note that further revision will be under the certificate no. LC20CA18552-1