Certification Record
LC1224503
C Technologies Inc.
2024-09-26
New Jersey
United States
IN-VPX-FLOW-A
29.260; 29.020
- Variable Pathlength Concentration Meter
- Model no.: IN-VPX-FLOW-A
- Ratings:
- HazLoc Ratings:
- Class I, Division 2, Groups A, B, C, and D T4 IP65
- Ambient temperature range: 0°C to 48°C.
- Electrical Ratings: 24VDC, 0.6A
- HazLoc Ratings:
- CSA-C22.2 NO. 213-17 (R2022) / UL 121201 (Ed.9)
- CAN/CSA-C22.2 NO. 61010-1-12/A1:18 (R2022)/UL61010-1 9Ed.3)
- To be supplied by a battery or Class 2 certified power supply if different from the ones assessed and described in the critical components list.
- The interconnection cables between different components to be provided by using power-limited tray cable (Type PLTC), as allowed for applications supplied by a Class 2 or Class 3 power source. Alternatively, all components shall be installed in a NEMA 4 enclosure.
- USB connection to be used only in non-hazardous areas.
- The power supply must be properly certified for the location or installed according to the CEC and NEC wiring methods accepted to the Local Authority Having Jurisdiction (LAHJ).
General Note:
- Project no.: T-22235A
- LC Listing report no.: 11.03.22235-1_Rev.A
- LC Certificate no.: LC22235-1