LC Clients

Certification Record

LC1235910
ABB Inc. (Real Tech Inc.)
2025-12-30
Ontario
Canada
Probe-BA-EX2010, Safety Barrier-ZB-12V2.3A
29.260.20
  • BOD/COD Probe
  • Model Nos.:
    • Probe – BA-EX2010
    • Safety Barrier – ZB-12V2.3A
  • Ratings:
    • Power:12Vdc, 2.3A
    • Hazardous Location Rating: Class I, Division 1, Groups C, D T4
    • Ta = 0oC to +40oC

Serial Nos.:

Model No. Serial No. LabTest Label Number
BA-EX2010 UVT-17-P1003407 14076
ZB-12V2.3A ZB-23-P11003407 14075

 

  • CSA C22.2 NO. 60079-0:19 (R2024)
  • CAN/CSA-C22.2 NO. 60079-11:14 (R2018)
  • CSA C22.2 NO. 60079-28:16 (R2021)
  • CAN/CSA-C22.2 NO.61010-1-12/A1:18 (R2022)
  • The maximum surface temperature rise has been assigned T4 (135°C). Any flammable substances with an auto-ignition temperature of less than 135°C must not be used.
  • No changes to or modifications of the intrinsically safe barriers are permitted without additional assessment.
  • Intrinsically safe cables from the safety barriers to the probe must be protected against mechanical damage during installation, operation and maintenance.
  • The cable length between the safety barrier in a non-hazardous area and probe must be less than 15 meters.
  • A separation of at least 50mm must be maintained between all the different intrinsically safe cable and non-intrinsically safe cables at all times.
  • Intrinsically safe when connected as per control drawing BA-EX_CONTROL_MAR_2023

General Note:

  • Project no.: 25-1164
  • LC Listing Report no.: 11.03.25-1164-1_Rev.0
  • LC Certificate no.: LC25-1164-1