LC Clients

Certification Record

LC1252748
Impossible Sensing Energy Inc.
2026-05-20
Alberta
Canada
Flow-MKV
29.260.20, 21.180 and 19.080
  • Multiphase Meter
  • Model: Flow-MKV
  • Ratings:
    • Hazardous Location Ratings:
      Class I, Division 2, Group D; T3
      Ambient temperature: -40°C to +40°C
      Process temperature: -40°C to +75°C
      Enclosure Type 4
    • Electric ratings: 24 Vdc, 1.7 A max
  • CSA C22.2 No. 94.2:20 (R2025): Enclosures for electrical equipment, environmental considerations (Tri-national standard with NMX-J-235/2-ANCE-2020 and UL 50E)
  • C22.2 No. 213-17 (R2022) – Nonincendive electrical equipment for use in Class I and II, Division 2 and Class III, Divisions 1 and 2 hazardous (classified) locations (bi-national with UL 121201)
  • CSA C22.2 No. 61010-1:2012/ UL 61010-1 3rd Edition: Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements
  • The equipment must be powered from a suitably certified Class 2 power supply.
  • The end user is to ensure appropriate earthing upon installation.
  • The final installation of the device shall be in full compliance with the applicable wiring method requirements of the Canadian Electrical Code (CEC) for Class I, Division 2 installations in Canada and the National Electrical Code (NEC) for installations in the United States and shall be subject to approval by the applicable Authorities Having Jurisdiction (AHJ).
  • End users shall follow specific installation and operating instructions supplied by the manufacturer to prevent exceeding the heat dissipating limits for the desired temperature code for an operating temperature ambient.
  • No changes to or modifications are permitted without additional assessment.

General Notes:

  • LC Project no.: 25-5721
  • LC Certificate no.: LC25-5721-1
  • LC Listing Report no.: 11.03.25-5721-1_Rev0
Go to Top