LC1234259
Canada Sensors Technology Inc.
2026-05-29
Alberta
Canada
Process 4; Process 4 HYD; Process 6; Process 6 HYD; Process 9 – IS; Process 9 - IS - HYD Pressure Transmitters
17.100 and 29.020
  • Pressure Transmitters
  • Model:  Process 4; Process 4 HYD; Process 6; Process 6 HYD; Process 9 – IS;
    Process 9 – IS – HYD Pressure Transmitters
  • Rated: Class I, Div. 2 Grp ABCD Class I Zone 2
    AEx ec IIC T4 Gc, Ex ec IIC T4 Gc
    IP65, Ta: -40°C to +95°C

    • Single Seal
      MWP: 10,000psi (Process 4 and Process 4 HYD), 20,000 psi (Process 6 and Process 6 HYD) and 200psi (Process 9 – IS and Process 9 – IS – HYD), 10-28Vdc, 5 mA
  • CSA C22.2 NO. 213-17 – Nonincendive electrical equipment for use in Class I and II, Division 2 and Class III, Divisions 1 and 2 hazardous (classified) locations
  • CSA-C22.2 NO. 61010-1-12 (R2017) – Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements
  • CSA-C22.2 NO. 60079-0:19 – Explosive atmospheres — Part 0: Equipment — General requirements
  • CSA-C22.2 NO. 60079-7A:16 – Explosive atmospheres — Part 7: Equipment protection by increased safety “e”
  • UL 121201 (Ed 9) – Nonincendive electrical equipment for use in Class I and II, Division 2 and Class III, Divisions 1 and 2 hazardous (classified) locations
  • UL 61010-1 (Ed 3) – Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements
  • UL 60079-0 (Ed 7) – Explosive atmospheres — Part 0: Equipment — General requirements
  • UL 60079-7 (Ed 5) – Explosive atmospheres — Part 7: Equipment protection by increase safety “e”
  • ANSI/ISA-12.27.01.2003 – Requirements for Process Sealing Between Electrical Systems and Flammable or Combustible Process Fluids
  • Installation to be permanently connected, Pollution Degree 4 or better, Overvoltage Category I.
  • Powered by certified Class 2 Power Supply as defined in the Canadian Electrical Code or National Electrical Code or by a CSA C22.2 No.60950 -1/ UL 60950-1 certified limited power source (LPS).
  • Wiring end must be installed in conduit or other enclosure providing IP65 ingress protection.
  • Models with DIN 43650A connector (nomenclature 22) are limited to Class I and not to be used for Class II applications.
  • Installation must include protection from impact greater than 4 joules.

General Note:

  • LC Project no.: 19352A
  • LC Certificate no.: LC26CA19352-1
  • LC Listing Report no.: 11.00.19352-1_RevA