Certification Record

LC1231537
EVTeam Corp.
2022-12-22
Ottawa
Canada
FLD-ABCL8-12V, FLD-ABCL8-18V, FLD-ABCL10-12V, FLD-ABCL10-18V, FLD-ABCL16-12V, FLD-ABCL16-18V, FLD-ERBC-12V, FLD-ERBC-18V
29.260; 29.260.20; 19.080; 71.040.10
  • Acquisition boards for interfacing with passive impedance sensors
  • Model Nos.: FLD-ABCL8-12V, FLD-ABCL8-18V, FLD-ABCL10-12V, FLD-ABCL10-18V,
    FLD-ABCL16-12V, FLD-ABCL16-18V, FLD-ERBC-12V, FLD-ERBC-18V
  • Ratings:
    • Associated Apparatus for use in Class I, Division 2, Groups A, B, C, and D; T6 Suitable for use in Class I, Zone 2, Group IIC – Provides intrinsically safe outputs for use in Class I, Division 1, Groups A, B, C, and D or Class I, Zone 0, Group IIC
    • Ta: -40°C to + 70°C
    • Input:18VDC Max, 50mA

Entity parameters:

Model No. Um Uo (V) Io (mA) Po (w) Co (uF) Lo (mH)
FLD-ABCL8-12V 250V 15 10 0.04 0.58 100
FLD-ABCL8-18V 22 14.7 0.081 0.165 100
FLD-ABCL10-12V 15 10 0.04 0.58 100
FLD-ABCL10-18V 22 14.7 0.081 0.165 100
FLD-ABCL16-12V 15 10 0.04 0.58 100
FLD-ABCL16-18V 22 14.7 0.081 0.165 100
FLD-ERBC-12V 15 147 0.55 0.58 1.65
FLD-ERBC-18V 22 147 0.81 0.165 1.65

 

 

  • CAN/CSA C22.2 No. 61010-1:2012
  • UL 61010-1 3rd Edition
  • CAN/CSA C22.2 No. 213: 2017
  • UL 121201: 2017
  • CAN/CSA C22.2 No. 60079-0:2019
  • CAN/CSA C22.2 No. 60079-11:2014
  • UL 60079-0:2019
  • UL 60079-11:2013
  • Acquisition boards shall be housed in an enclosure acceptable to the local authorities having jurisdiction.
  • The final enclosure must bear the following warning: “Do not open when an explosive atmosphere is present”.
  • The final installation of the device shall meet the requirements of CEC (for Canada) and NEC (for USA) for wiring method in Zone 2 and Division 2 and is subjected to acceptance of local authorities having jurisdiction.

General Note

  • Project no.: 20280A
  • LC Listing Report no.: 11.03.20280-1 _Rev.A
  • LC Certificate no.: LC20280-1